13 Symbol indication / Troubleshooting
Display symbol Condition/Cause Corrective action
The flashing heart will appear when the
pulse is detected during a measurement.
Measurements in process.
Do not talk or move.
Appears when the battery voltage is
excessively low or the position of batteries
is incorrect.
Replace all batteries with new ones.
Make sure the +/- polarity is correct.
Appears for 1 minute when the user was
talking, moving or shaking or an irregular
heart beat was detected during
measurement.
Repeat the measurement.
Note that you be relaxed for at least 5
minutes and sit comfortably and
quietly before you restart a
measurement.
Indicates the average of last
3 measurements.
Indicates the number of measurements
stored in the memory.
Appears when measurement error occurs
or the blood pressure value is displayed
excessively low or high.
Replace all batteries with new ones.
Make sure the +/- polarity is correct.
Air circuit abnormality.
Cuff tube may not be plugged into monitor
correctly.
Check the cuff connection.
Measure again.
Inflation pressure exceeding 300 mmHg. Turn the device off, then measure
again.
Error determining measurement data. Measure again.